کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
2939425 1405516 2016 8 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Overview of the 2016 U.S. Food and Drug Administration Circulatory System Devices Advisory Panel Meeting on the Absorb Bioresorbable Vascular Scaffold System
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی کاردیولوژی و پزشکی قلب و عروق
پیش نمایش صفحه اول مقاله
Overview of the 2016 U.S. Food and Drug Administration Circulatory System Devices Advisory Panel Meeting on the Absorb Bioresorbable Vascular Scaffold System
چکیده انگلیسی

ObjectivesThis study aims to describe the discussions and recommendations made during the U.S. Food and Drug Administration (FDA) Circulatory System Device Panel pre-market approval application for the Absorb Bioresorbable Vascular Scaffold (BVS) System.BackgroundThe Absorb BVS System is a first-of-its-kind fully bioresorbable percutaneous coronary intervention technology. The absorb BVS was studied in the ABSORB III (A Clinical Evaluation of Absorb BVS, the Everolimus Eluting Bioresorbable Vascular Scaffold in the Treatment of Subjects with de Novo Native Coronary Artery Lesions) trial, the pivotal U.S. investigational device exemption trial.MethodsObservational report of the FDA Circulatory System Device Panel pre-market approval application meeting held on March 15, 2016.ResultsThe U.S. FDA Circulatory System Device Panel members reviewed the ABSROB III trial outcomes and additional post hoc analyses presented by the sponsor and the FDA. The ABSORB III trial met the primary endpoint of noninferiority of Absorb BVS compared with the control, XIENCE drug-eluting stent, for target lesion failure at 1 year. Although a higher numerical trend for adverse outcomes was reported for the Absorb BVS, there were no statistical differences between Absorb BVS and XIENCE for any safety or effectiveness components for target lesion failure or for the secondary pre-specified outcomes. Panel members raised concerns with regard to the ABSORB III results and post hoc analyses focusing mainly on the noninferiority design of the trial, the apparent safety issues of the Absorb BVS in small vessels, the mismatch of visually versus intravascular imaging assessed vessel size found in ABSORB III and its implications on the adequate device labeling, the safety of Absorb BVS in specific patient and lesion subsets, and the post-approval commitments of the sponsor.ConclusionsFollowing panel discussions and the evidence presented, the panel voted for approval of the device.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: JACC: Cardiovascular Interventions - Volume 9, Issue 17, 12 September 2016, Pages 1757–1764
نویسندگان
, , , ,