Agreement of spatio-temporal gait parameters between a vertical ground reaction force decomposition algorithm and a motion capture system
ترجمه فارسی عنوان
توافق پارامترهای پارامتر زمانی و زمانی بین الگوریتم تجزیه ی نیروی واکنش زمین عمودی و یک سیستم ضبط حرکت
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
چکیده انگلیسی


• An algorithm that decomposes ground reaction forces and estimates spatiotemporal gait parameters is suggested.
• The algorithm allowed accurate and precise spatiotemporal gait parameters.
• Algorithm's gait parameters are in agreement with those of the motion capture system.
• Gait parameters computed with the algorithm are highly reproducible.

IntroductionA ground reaction force decomposition algorithm based on large force platform measurements has recently been developed to analyze ground reaction forces under each foot during the double support phase of gait. However, its accuracy for the measurement of the spatiotemporal gait parameters remains to be established.ObjectiveThe aim of the present study was to establish the agreement between the spatiotemporal gait parameters obtained using (1) a walkway (composed of six large force platforms) and the newly developed algorithm, and (2) an optoelectronic motion capture system.MethodsTwenty healthy children and adolescents (age range: 6–17 years) and 19 healthy adults (age range: 19–51 years) participated in this study. They were asked to walk at their preferred speed and at a speed that was faster than the preferred one. Each participant performed three blocks of three trials in each of the two walking speed conditions.ResultsThe spatiotemporal gait parameters measured with the algorithm did not differ by more than 2.5% from those obtained with the motion capture system. The limits of agreement represented between 3% and 8% of the average spatiotemporal gait parameters. Repeatability of the algorithm was slightly higher than that of the motion capture system as the coefficient of variations ranged from 2.5% to 6%, and from 1.5% to 3.5% for the algorithm and the motion capture system, respectively.ConclusionThe proposed algorithm provides valid and repeatable spatiotemporal gait parameter measurements and offers a promising tool for clinical gait analysis. Further studies are warranted to test the algorithm in people with impaired gait.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Gait & Posture - Volume 43, January 2016, Pages 257–264
نویسندگان
, , , , ,