Is Grey Level a Suitable Alternative to Low-Contrast Penetration as a Serial Measure of Sensitivity in Computerised Ultrasound Quality Assurance?
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم آکوستیک و فرا صوت
پیش نمایش صفحه اول مقاله
Is Grey Level a Suitable Alternative to Low-Contrast Penetration as a Serial Measure of Sensitivity in Computerised Ultrasound Quality Assurance?
چکیده انگلیسی
The aim of this study was to test the hypothesis that grey levels are a suitable alternative measure of sensitivity in ultrasound imaging quality assurance, as there are several caveats in the use of penetration depth. In a primary cohort of nine probes, where measurements had been made for 6 to 34 mo, both penetration depth and mean grey level fell below tolerance for six probes; both penetration depth and mean grey level remained within tolerance for three probes. In a secondary cohort where a measurement programme had been in place for a shorter period, grey level and/or penetration depth fell below tolerance in 15 of 66 probes; the sensitivity and specificity of at least 10% loss of grey level in predicting >5% loss in penetration depth were 91% and 93%, respectively. A loss of grey level accompanies a loss of penetration and provides a suitable alternative measure of sensitivity.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Ultrasound in Medicine & Biology - Volume 43, Issue 2, February 2017, Pages 541-545