کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
5499091 1533484 2017 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Medical and occupational dose reduction in pediatric barium meal procedures
ترجمه فارسی عنوان
کاهش دوزهای پزشکی و شغلی در روش غذا باریم کودکان
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
چکیده انگلیسی


- Technique optimization was made following the EC and ICRP recommendations.
- Optimization reduced 60-70% patient dose.
- Occupational exposure was reduced about 50%.

Doses received in pediatric Barium Meal procedure can be rather high. It is possible to reduce dose values following the recommendations of the European Communities (EC) and the International Commission on Radiological Protection (ICRP). In the present work, the modifications of radiographic techniques made in a Brazilian hospital according to the EC and the ICRP recommendations and their influence on medical and occupational exposure are reported. The procedures of 49 patients before and 44 after the optimization were studied and air kerma-area product (PK,A) values and the effective doses were evaluated. The occupational equivalent doses were measured next to the eyes, under the thyroid shield and on each hand of both professionals who remained inside the examination room. The implemented modifications reduced by 70% and 60% the PK,A and the patient effective dose, respectively. The obtained dose values are lower than approximately 75% of the results from similar studies. The occupational annual equivalent doses for all studied organs became lower than the limits set by the ICRP. The equivalent doses in one examination were on average below than 75% of similar studies.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Radiation Physics and Chemistry - Volume 140, November 2017, Pages 271-274
نویسندگان
, , , , , , , , , ,