کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
737626 1461912 2012 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Evaluation of the packaging and encapsulation reliability in fully integrated, fully wireless 100 channel Utah Slant Electrode Array (USEA): Implications for long term functionality
موضوعات مرتبط
مهندسی و علوم پایه شیمی الکتروشیمی
پیش نمایش صفحه اول مقاله
Evaluation of the packaging and encapsulation reliability in fully integrated, fully wireless 100 channel Utah Slant Electrode Array (USEA): Implications for long term functionality
چکیده انگلیسی

The encapsulation and packaging reliability in fully integrated, fully wireless 100 channel Utah Slant Electrode Array (USEA)/integrated neural interface-recording version 5 (INI-R5) has been evaluated by monitoring the extended long term in vitro functional stability and recording longevity. The INI encapsulated with 6-μm Parylene-C was immersed in phosphate buffer saline (PBS) at room temperature for a period of over 12 months. The USEA/INI-R5, while being soaked was powered and configured wirelessly through 2.765 MHz inductive link and the transmitted frequency shift keying (FSK) modulated radio-frequency (RF) (900 MHz industrial, scientific, medical-ISM band) signal was also recorded wirelessly as a function of soak time. In order to test the long term recording ability, in vitro wireless recording was performed in agarose for few channels. The full functionality and the ability of the electrodes to record artificial neural signals even after 12 months of PBS soak provides a measure of encapsulation reliability, the functional and recording stability in fully integrated wireless neural interface and potential usefulness for future chronic implants.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Sensors and Actuators A: Physical - Volume 188, December 2012, Pages 167–172
نویسندگان
, , , , , , , , , , ,