Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
10081732 | The Journal of Arthroplasty | 2005 | 8 Pages |
Abstract
This study was designed to elucidate the erosive effect of hydrogen peroxide solutions on the materials used for total-hip arthroplasty (THA). As test materials, cross-linked polyethylene, Ti-6Al-4V alloy, and thermal sprayed hydroxyapatite (HA) were used. Changes upon soaking in 3% hydrogen peroxide, before soaking, 1 minute after soaking, 10 minutes after soaking, and 180 minutes after soaking were examined. Scanning electron microscope, Fourier transform infrared analysis, and x-ray diffraction were used for this examination. Hydrogen peroxide did not affect polyethylene, although notable changes in the Ti-6Al-4V alloy and HA did occur. These results indicate that caution should also be exercised to minimize erosion of prosthesis consisting of HA and Ti alloy when hydrogen peroxide solutions are used during total-hip arthroplasty.
Related Topics
Health Sciences
Medicine and Dentistry
Orthopedics, Sports Medicine and Rehabilitation
Authors
Masamori MD, Masaru MD, Kenji MD, Takami MD, Takao MD,