Article ID Journal Published Year Pages File Type
10136966 Applied Radiation and Isotopes 2018 6 Pages PDF
Abstract
The dosimetric response of a multi guard ring structure (MGR) diode has been studied with clinical electron beam energies from 5 MeV to 15 MeV. The results showed that the MGR dose response is linear in the range of 5-320 cGy and presents reproducibility with variation coefficients less than 0.4%. The field output factors measured with the MGR agreed within 2% with those measured with an ionization chamber. This study evidences that this diode can be used for clinical electron beam dosimetry.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation
Authors
, , , , , , ,