Abstract
Two new polymer-based stationary phases; DCpak PBT (poly(butylene terephthalate)) and DCpak P4VP (poly(4-vinylpirydine)) were evaluated for the analysis of polyphenols using supercritical fluid chromatography (SFC). The compounds studied included phenolic acids and flavonoids. The different variables that influence the chromatographic separation, such as type and percentage of organic modifier, additive, pressure and temperature were examined. Using the DCpak P4VP column the retention was exceptionally high, obtaining better results with the DCpak PBT column. The separation of nine polyphenols was achieved using a gradient of modifier (methanol with 0.1% trifluoroacetic acid) from 5 to 50%, a pressure of 150â¯bar, a temperature of 35â¯Â°C and a flow-rate of 2â¯mL/min. The use of additives was necessary in order to obtain good peak shapes and efficiencies, achieving the best results with trifluoroacetic acid. LODs and LOQs values were lower than 5â¯Î¼g/mL in all the cases; meanwhile, the %RSD values for method repeatability and inter-day reproducibility were lower than 3% and 10% respectively. Finally, the proposed method was successfully applied to the analysis of polyphenols in commercial bee pollen; four compounds, namely cinnamic acid, p-coumaric acid, catechin and quercetin were identified and quantified.