Abstract
The intermediate radiological results with the AcSys Shearer Cup were unsatisfactory because of the high loosening rate, although the revision rate was low. The non-porous outer surface and the poor fixation mechanism between the metal shell and liner may have contributed to the high failure rate. Regular radiological review is recommended when this cup is used because early loosening is often painless.
Related Topics
Health Sciences
Medicine and Dentistry
Orthopedics, Sports Medicine and Rehabilitation
Authors
Masahide Ikema, Masamizu Oyama, Atsushi Kita, Kanichi Funayama,