Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
10407647 | Measurement | 2005 | 8 Pages |
Abstract
The multi-channel thermometer was calibrated by comparison against a platinum resistance thermometer traceable to ITS-90 in the temperature range 0-150 °C. The system affords a temperature repeatability to within 0.3 °C, with 0.5 s response time. Overall system performance and probe calibration are reported. Examples of temperature distribution measurements within food analogues heated in microwave ovens are reported, as well.
Related Topics
Physical Sciences and Engineering
Engineering
Control and Systems Engineering
Authors
V.C. Fernicola, R. Galleano,