Article ID Journal Published Year Pages File Type
10407647 Measurement 2005 8 Pages PDF
Abstract
The multi-channel thermometer was calibrated by comparison against a platinum resistance thermometer traceable to ITS-90 in the temperature range 0-150 °C. The system affords a temperature repeatability to within 0.3 °C, with 0.5 s response time. Overall system performance and probe calibration are reported. Examples of temperature distribution measurements within food analogues heated in microwave ovens are reported, as well.
Related Topics
Physical Sciences and Engineering Engineering Control and Systems Engineering
Authors
, ,