| Article ID | Journal | Published Year | Pages | File Type | 
|---|---|---|---|---|
| 10409319 | Sensors and Actuators A: Physical | 2011 | 8 Pages | 
Abstract
												We demonstrate the operation of a digital microfluidic lab-on-a-chip system utilizing Electro Wetting on Dielectrics (EWOD) as the actuation principle and a High Fundamental Frequency (HFF; 50 MHz) quartz crystal microbalance (QCM) resonator as a mass-sensitive sensor. In a first experiment we have tested the reversible formation of a phosphor-lipid monolayer of phospholipid vesicles out of an aqueous buffer suspension onto a bio-functionalized integrated QCM sensor. A binding of bio-molecules results in an altered mass load of the resonant sensor and a shift of the resonance frequency can be measured. In the second part of the experiment, the formation of a protein multilayer composed of the biomolecule streptavidin and biotinylated immunoglobulin G was monitored. Additionally, the macroscopic contact angle was optically measured in order to verify the bio-specific binding and to test the implications onto the balance of the surface tensions. Using these sample applications, we were able to demonstrate and to verify the feasibility of integrating a mass-sensitive QCM sensor into a digital microfluidic chip.
											Related Topics
												
													Physical Sciences and Engineering
													Chemistry
													Electrochemistry
												
											Authors
												Thomas Lederer, Brigitte P. Stehrer, Siegfried Bauer, Bernhard Jakoby, Wolfgang Hilber, 
											