Article ID Journal Published Year Pages File Type
10429677 Biosensors and Bioelectronics 2005 7 Pages PDF
Abstract
This paper presents disposable protein analysis chips with single- or four-chamber-constructed from poly(dimethylsiloxane) (PDMS) and silicon. The chips are composed of a multilayer stack of PDMS layers that sandwich a silicon microchip. This inner silicon chip features an etched array of microcavities hosting polymeric beads. The sample is introduced into the fluid network through the top PDMS layer, where it is directed to the bead chamber. After reaction of the analyte with the probe beads, the signal generated on the beads is captured with a CCD camera, digitally processed, and analyzed. An established bead-based fluorescent assay for C-reactive protein (CRP) was used here to characterize these hybrid chips. The detection limit of the single-chamber protein chip was found to be 1 ng/ml. Additionally, using a back pressure compensation method, the signals from each chamber of the four-chamber chip were found to fall within 10% of each other.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , , , , , , , ,