Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
10430235 | Biosensors and Bioelectronics | 2005 | 9 Pages |
Abstract
A novel holographic sensor system capable of detecting dynamic changes in glucose concentration has been developed. The hologram is recorded within a bio-compatible hydrogel matrix containing phenylboronic acid derivatives. On binding glucose, the colour of the hologram red-shifts to longer wavelengths as the hydrogel expands and this colour change is used to quantify glucose concentration. However, phenylboronic acids are non-selective and bind a wide variety of cis-diols. In blood, glucose is the only sugar found free at high concentration, whilst other sugars are typically found as part of glycoproteins and macromolecular structures. Although glycoproteins have been shown to have no effect on the sensor, phenylboronic acids can bind lactate much more readily than glucose. We have designed two polymer hydrogel systems to increase the selectivity of the sensor for glucose over lactate. The first involved the use of high concentrations of 3-acrylamidophenylboronic acid (3-APB) whilst the second system utilised 2-acrylamido-5-fluorophenylboronic acid (5-F-2-MAPB). Both systems displayed an increased selectivity to glucose over lactate at physiological pH and ionic strength and could be deployed as selective holographic sensors for glucose detection in physiological fluids.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Satyamoorthy Kabilan, Alexander J. Marshall, Felicity K. Sartain, Mei-Ching Lee, Abid Hussain, Xiaoping Yang, Jeff Blyth, Njeri Karangu, Karen James, Jimmy Zeng, Dawn Smith, Angelika Domschke, Christopher R. Lowe,