Article ID Journal Published Year Pages File Type
10432837 Journal of Biomechanics 2011 6 Pages PDF
Abstract
The present study found that, with an increase in stenosis severity, FFR decreased whereas CDP and LFC increased. The cutoff value of 0.75 for FFR was observed at 78.7% area stenosis for RR, whereas for CC and CS the cutoff values were obtained at higher stenosis severities of 81.3% and 82.7%, respectively. For a fixed stenosis, CDP value decreased and LFC value increased with a decrease in plaque elasticity (RR to CS). We conclude that the differences in diagnostic parameters with compliance at intermediate stenosis (78.7-82.7% area blockage) could lead to misinterpretation of the stenosis severity.
Related Topics
Physical Sciences and Engineering Engineering Biomedical Engineering
Authors
, , ,