Article ID Journal Published Year Pages File Type
10434527 Journal of Biomechanics 2005 7 Pages PDF
Abstract
In the present study, 32 patients were implanted with a mobile-bearing LCS TKA utilizing the balanced gap technique. An electronic pressure sensor, developed specifically to record pressure magnitude and distribution in the medial and lateral compartments, was incorporated into the implant trials. The knee was then passively taken through a range of motion while pressure data was recorded via computer. Postoperatively, 16 patients underwent active fluoroscopic kinematic analysis to assess for condylar liftoff and femorotibial translation. We found that abnormal compartment pressures and distributions as recorded by the intraoperative pressure sensor were correlated with inappropriate or paradoxical postoperative kinematics. In addition, subjects having similar pressures in both compartments throughout a range of motion did not experience condylar liftoff values greater than 1.0 mm. These data suggest that surgical technique influences the magnitude and distribution of forces at the articulation, postoperative kinematics, and likely, implant longevity.
Related Topics
Physical Sciences and Engineering Engineering Biomedical Engineering
Authors
, , ,