Article ID Journal Published Year Pages File Type
10435141 Medical Engineering & Physics 2005 10 Pages PDF
Abstract
The FE model of the posterolateral fusion with the screw showed that screws on the caudal side had larger axial stress than those on the cephalic side, supporting the clinical findings that 75% of the patients had the screw breakage on the caudal side. SEM fractography showed that all broken screws exhibited beach marks or striations on the fractured surface, indicating fatigue failure. Screws of patients with spinal fracture showed fatigue striations and final ductile fracture around the edge. Among the 16 patients who had broken pedicle screws 69% of them achieved bone union in the bone graft, showing that bone union in the bone graft did not warrant the prevention of screw breakage.
Related Topics
Physical Sciences and Engineering Engineering Biomedical Engineering
Authors
, , , , , ,