Article ID Journal Published Year Pages File Type
10542416 Food Chemistry 2013 6 Pages PDF
Abstract
A new method was developed for the determination of organic and inorganic selenium in selenium-enriched rice by graphite furnace atomic absorption spectrometry detection after cloud point extraction. Effective separation of organic and inorganic selenium in selenium-enriched rice was achieved by sequentially extracting with water and cyclohexane. Under the optimised conditions, the limit of detection (LOD) was 0.08 μg L−1, the relative standard deviation (RSD) was 2.1% (c = 10.0 μg L−1, n = 11), and the enrichment factor for selenium was 82. Recoveries of inorganic selenium in the selenium-enriched rice samples were between 90.3% and 106.0%. The proposed method was successfully applied for the determination of organic and inorganic selenium as well as total selenium in selenium-enriched rice.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , ,