Article ID Journal Published Year Pages File Type
10546344 Journal of the American Society for Mass Spectrometry 2005 10 Pages PDF
Abstract
The feasibility of ultra thin-layer chromatography atmospheric pressure matrix-assisted laser desorption ionization mass spectrometry (UTLC-AP-MALDI-MS) has been studied in the analysis of small molecules. Because of a thinner adsorbent layer, the monolithic UTLC plates provide 10-100 times better sensitivity in MALDI analysis than conventional high performance thin-layer chromatography (HPTLC) plates. The limits of detection down to a low picomole range are demonstrated by UTLC-AP-MALDI-MS. Other advantages of UTLC over HPTLC include faster separations and lower solvent consumption. The performances of AP-MALDI-MS and vacuum MALDI-MS have been compared in the analysis of small drug molecules directly from the UTLC plates. The desorption from the irregular surface of UTLC plates with an external AP-MALDI ion source combined with an ion trap instrument provides clearly less variation in measurements of m/z values when compared with a vacuum MALDI-time-of-flight (TOF) instrument. The performance of the UTLC-AP-MALDI-MS method has been applied successfully to the purity analysis of synthesis products produced by solid-phase parallel synthesis method.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , , , ,