Article ID Journal Published Year Pages File Type
10546677 Journal of the American Society for Mass Spectrometry 2005 5 Pages PDF
Abstract
A minimum thickness gate valve design for mass spectrometry is described in detail. The ion optics required to transmit ions from the source to the ICR cell are integrated into the design to minimize fringe field effects on the ions as they travel through the gate valve. The total thickness of the complete gate valve assembly is 1.03 in. (26.2 mm) with a maximum fringe field distance of 0.065 in. (1.7 mm). The gate valve is able to maintain a vacuum of <10−10 mbar at the ICR cell when the source is vented to atmosphere and the estimated ion transfer efficiency is >95%.