Article ID Journal Published Year Pages File Type
10546957 Journal of the American Society for Mass Spectrometry 2005 10 Pages PDF
Abstract
Bioearosol mass spectrometry (BAMS) analyzes single particles in real time from ambient air, placing strict demands on instrument sensitivity. Modeling of the BAMS reflectron time of flight (TOF) with SIMION revealed design limitations associated with ion transmission and instrument sensitivity at higher masses. Design and implementation of a BAMS linear TOF with electrostatic ion guide and delayed extraction capabilities has greatly increased the sensitivity and mass range relative to the reflectron design. Initial experimental assessment of the new instrument design revealed improved sensitivity at high masses as illustrated when using standard particles of cytochrome C (m/z ∼ 12,000), from which the compound's monomer, dimer (m/z ∼ 24,000) and trimer (m/z ∼ 36,000) were readily detected.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , , , , ,