Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
10548235 | Journal of Chromatography A | 2005 | 7 Pages |
Abstract
A method was developed using dynamic headspace liquid-phase microextraction and gas chromatography-mass spectrometry for extraction and determination of 9 alcohols from water samples. Four different solvents, hexyl acetate, n-octanol, o-xylene and n-decane were studied as extractants. The analytes were extracted using 0.8 μl of n-octanol from the headspace of a 2 ml sample solution. The effect of sampling volume, solvent volume, sample temperature, syringe plunger withdrawal rate and ionic strength of the solution on the extraction performance were studied. A semiautomated system including a variable speed stirring motor was used to ensure a uniform movement of syringe plunger through the barrel. The method provided a fairly good precision for all compounds (5.5-9.3%), except methanol (16.4%). Detection limits were found to be between 1 and 97 μg/l within an extraction time of â¼9.5 min under GC-MS in full scan mode.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Mohammad Saraji,