Article ID Journal Published Year Pages File Type
10550086 Journal of Chromatography B 2005 4 Pages PDF
Abstract
A simple method, exposure to natural-light, was developed to remove riboflavin from urine to enhance its use as the biological matrix for the preparation of calibration and control samples. Riboflavin-depleted urine containing less than 1 ng/ml of riboflavin was used to validate a high-performance liquid chromatography with fluorescence detection method for the determination of urinary riboflavin. The linearity of the assay (r2 = 0.999) was acceptable over the range of 10-5000 ng/ml. The intra-assay and inter-assay CVs were 3.3% and 9%, respectively. Subsequent stability studies found that urine riboflavin was stable for up to 6 months at 4 or −20 °C.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry