A sensitive and reproducible method is described for the analysis of trichloroacetic acid in urine and 1,1,1-trichloroethane in blood using dynamic headspace GC/MS. Samples were analyzed using the soil module of a modified purge and trap autosampler to facilitate the use of disposable purging vessels. Coefficients of variation were below 3.5% for both analytes, and response was linear in the range of 0.01-7.0 μg/ml for trichloroacetic acid and 0.9 ng/ml-2.2 μg/ml for 1,1,1-trichloroethane. Attempts at using dynamic headspace for the analysis of trichloroethanol in urine were unsuccessful.