Article ID Journal Published Year Pages File Type
10550142 Journal of Chromatography B 2005 7 Pages PDF
Abstract
Capillary electrophoresis (CE) has become a useful analytical tool for the analysis of microdialysis samples. However, CE with UV detection (CE-UV) does not provide detection limits sufficient to quantify glutathione (GSH) and glutathione disulfide (GSSG) in biological samples such as liver microdialysates, because of the small optical path length in the capillary. To overcome this limitation, an on-column preconcentration technique, pH-mediated base stacking, was used in this study to improve the sensitivity of CE-UV. This stacking technique allowed large volumes of high ionic strength sample injection without deterioration of the separation efficiency and resolution. A 26-fold increase in sensitivity was achieved for both GSH and GSSG using the pH-mediated base stacking, relative to normal injection without stacking. The limit of detection for GSH and GSSG was found to be 0.75 μM (S/N = 6) and 0.25 μM (S/N = 6), respectively. The developed method was used to analyze GSH and GSSG in liver microdialysates of anesthetized Sprague Dawley male rats. The basal concentrations of GSH and GSSG in the liver microdialysates of male rats were found to be 4.73 ± 2.08 μM (n = 7) and 5.52 ± 3.66 μM (n = 7), respectively.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , ,