Article ID Journal Published Year Pages File Type
10550180 Journal of Chromatography B 2005 7 Pages PDF
Abstract
High performance liquid chromatographic (HPLC) methods were validated for the determination of aripiprazole (OPC-14597, Abilify™) in rat plasma and brain. Separation was by Nova-pak phenyl column; flow rate, 1.0 ml/min; mobile phase, acetonitrile-methanol-20 mM sodium sulfate-acetic acid (27:25:48:1, v/v/v/v); UV detection at 254 nm. Reproducibility in plasma and brain showed excellent precision (within 7.8 and 10.6%) and accuracy (96.0-102.4% and 99.0-108.7%) with calibration curve ranges 10.0-2000 ng/ml and 30.0-6000 ng/g, respectively. Validated HPLC methods were successfully applied to pharmacokinetic study of aripiprazole in rats, demonstrating brain concentrations after oral administration five times higher than plasma concentrations.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , ,