Article ID Journal Published Year Pages File Type
10550220 Journal of Chromatography B 2005 5 Pages PDF
Abstract
A robust, rapid, selective and sensitive liquid chromatography-negative atmospheric pressure chemical ionization (LC-(APCI−)-MS-MS) method has been developed for the quantification of mometasone furoate (MF) in human plasma utilizing a solid-phase extraction clean-up step and 13C-fluticasone propionate as internal standard. The intra- and inter-day coefficients of variation were ≤15% and the lower limit of quantification (LLOQ) was 15 pg/ml. This method is ideally suited for pharmacokinetic investigations of low MF levels following inhalation of MF.