Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
10551178 | Journal of Chromatography B | 2005 | 5 Pages |
Abstract
Amosulalol is an antihypertensive drug with selective postsynaptic alpha 1 and non-selective beta blocking effects. A simple solid-phase extraction and high-performance liquid chromatographic (HPLC) method has been developed and validated for the quantitative determination of amosulalol in human plasma. A reversed phase C18 column was used for the separation of amosulalol and ethyl paraben (internal standard) with a mobile phase composed of 0.025 M phosphate buffer (pH 6.0)·acetonitrile (73:27, v/v) at a flow rate of 1.5 mL/min. The ultraviolet detector was operated at the 272 nm wavelength. Intra- and inter-day precision and accuracy were acceptable for all quality control samples including the lower limit of quantification of 30 ng/mL. Recovery of amosulalol from human plasma was >95.6%. Amosulalol was stable in human plasma under various storage conditions. This method was used successfully for a pharmacokinetic study in plasma after oral administration of a single 20 mg dose of amosulalol hydrochloride to 16 healthy volunteers.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Hye Sun Gwak, Eun Sun Noh, Nah Young Lee, In Koo Chun,