Article ID Journal Published Year Pages File Type
10552218 Journal of Chromatography B 2005 9 Pages PDF
Abstract
A derivatization reagent, 2-hydrazino-1-methylpyridine, was developed for the liquid chromatography-electrospray ionization-mass spectrometry (LC-ESI-MS) of oxosteroids. The reagent quantitatively reacted with oxosteroids at 60 °C within 1 h and the resulting derivatives of the mono-oxosteroids provided a 70-1600-fold higher sensitivity compared to intact steroids. However, HMP was unsuitable for di-oxosteroids, such as androstenedione and progesterone. The developed derivatization procedure was applied to the LC-ESI-MS analysis of 5α-dihydrotestosterone in human prostate, and allowed the reproducible quantification of nanogram/gram level of the androgen with a 10-mg sample.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , ,