Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
10553586 | Journal of Pharmaceutical and Biomedical Analysis | 2011 | 10 Pages |
Abstract
The whole methodology was fully validated according to UNI EN ISO/IEC 17025:2005 rules and 2006 SOFT/AAFS guidelines. Selectivity, linearity range, identification (LOD) and quantitation (LOQ) limits, precision, accuracy and recovery were evaluated. For all the species the signal/concentration linearity was satisfactory in the 50-1000 ng/mL concentration range. The limits of detection ranged from 0.5 to 30 ng/mL and LOQs from 1.7 to 100.0 ng/mL. Precisions were in the ranges 5.0-11.8%, 1.5-11.0% and 1.1-4.4% for low (100 ng/mL), medium (300 ng/mL) and high (1000 ng/mL) concentration, respectively. The accuracy, expressed as bias% was within ±25% for all the analytes. The recovery values, evaluated at 300 ng/mL concentration, ranged from 56.2% to 98.8%. The present method for the determination of several benzodiazepines, zolpidem and zopiclone in human urine proved to be simple, fast, specific and sensitive. The quantification by LC-MS/MS was successfully applied to 329 forensic cases among driving re-licensing, car accidents and alleged sexual violence cases.