Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
10553692 | Journal of Pharmaceutical and Biomedical Analysis | 2005 | 8 Pages |
Abstract
This paper describes the development of a gas chromatography (GC) method used for the assay of isotretinoin in its isolated form and in pharmaceutical formulations. Isotretinoin soft and hard gelatin capsules were prepared with various excipients. The performance of the proposed gas chromatography method was compared to that of traditional high performance liquid chromatography (HPLC) systems for this substance, and the GC parameters were established based on several preliminary tests, including thermal analysis of isotretinoin. Results showed that gas chromatography-flame ionization detector (GC-FID) exhibited a separation efficiency superior to that of HPLC, particularly for separating isotretinoin degradation products. This method was proven to be effectively applicable to stability evaluation assays of isotretinoin and isotretinoin based pharmaceuticals.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Eliana Martins Lima, Danielle G. Almeida Diniz, Nelson R. Antoniosi-Filho,