Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
10559674 | Talanta | 2011 | 10 Pages |
Abstract
The optimum assay conditions were achieved with a Chromolith⢠Performance RP-8e 100 mm Ã 4.6 mm column and a mixture of methanol/acetonitrile/triethylamine/diammonium hydrogen phosphate (pH 6.25, 0.048 mol Lâ1) (20:20:0.1:59.9, v/v/v/v) as the mobile phase at a flow rate of 2 mL minâ1 and detection wavelength of 261 nm. The run time was less than 5 min, which is much shorter than the previously optimized methods. The optimized method was validated according to FDA guidelines to confirm specificity, linearity, accuracy and precision.