Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
10559759 | Talanta | 2011 | 10 Pages |
Abstract
This work details a rapid analytical method using direct sample injection for the simultaneous identification/quantification of 22 drugs of abuse, including some of their major metabolites, in environmental samples. This has been developed using a hybrid triple quadrupole-linear ion trap-mass spectrometer (QqLIT). With the increasing sensitivity of today's tandem mass spectrometers, direct injection analysis of water samples has become an attractive alternative to traditional analytical protocols, which often include a preliminary pre-concentration step. What's more, this kind of analysis is in accordance with many of the main objectives of so-called green analytical chemistry, or environmentally friendly practice. The analytical performance of the LC-MS/MS method was evaluated in three different water matrices (surface water, influent and effluent wastewater). Data acquisition was carried out in selected reaction monitoring (SRM) mode under time-scheduled conditions, monitoring two SRM transitions for simultaneous identification/quantification of all target compounds in the samples. Additionally, an experiment was performed using the information-dependent acquisition (IDA) scan to carry out the identification of those analytes for which the second transition was present at a low intensity. Finally, the two methodologies developed were applied to real samples for evaluation.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
M.J. MartÃnez Bueno, S. Uclés, M.D. Hernando, A.R. Fernández-Alba,