Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
10559795 | Talanta | 2011 | 6 Pages |
Abstract
This study deals with O2â generation in corona discharge (CD) in point to plane geometry for single flow ion mobility spectrometry (IMS) with gas outlet located behind the ionization source. We have designed CD of special geometry in order to achieve the high O2â yield. Using this ion source we have achieved in zero air conditions that up to 74% all negative ions were O2â or O2â(H2O). It has been demonstrated that the non-electronegative nitrogen positively influences the efficiency of O2â generation in O2/N2 mixtures. The reduced ion mobility of 2.27Â cm2Â Vâ1Â sâ1 has been measured for O2â/O2â(H2O) ions in zero air. Additional ions detected in zero air (less than 200Â ppb CO2) using the mass spectrometric and IMS technique were, NO2â, N2O2â (2.37Â cm2Â Vâ1Â sâ1), NO3â, N2O3â and N2O3â(H2O). The CO3â and CO4â ions have been detected after the introduction of 5Â ppm CO2 into zero air.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Martin Sabo, Ján MatúÅ¡ka, Å tefan MatejÄÃk,