Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
10559906 | Talanta | 2011 | 5 Pages |
Abstract
An automatic sequential injection system, combining monosegmented flow analysis, sequential injection analysis and sequential injection titration is proposed for acidity determination. The system enables controllable sample dilution and generation of standards of required concentration in a monosegmented sequential injection manner, sequential injection titration of the prepared solutions, data collecting, and handling. It has been tested on spectrophotometric determination of acetic, citric and phosphoric acids with sodium hydroxide used as a titrant and phenolphthalein or thymolphthalein (in the case of phosphoric acid determination) as indicators. Accuracy better than |4.4|% (RE) and repeatability better than 2.9% (RSD) have been obtained. It has been applied to the determination of total acidity in vinegars and various soft drinks. The system provides low sample (less than 0.3Â mL) consumption. On average, analysis of a sample takes several minutes.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Joanna Kozak, Marzena Wójtowicz, Nadzieja Gawenda, PaweÅ KoÅcielniak,