Abstract
												A method for the sequential enantiomeric quantification of d-malate and l-malate by a single line flow-injection analysis was developed using immobilized-enzyme reactors and fluorescence detection. An immobilized d-malate dehydrogenase (d-MDH) reactor and an immobilized l-malate dehydrogenase (l-MDH) reactor were introduced into the flow line in series. Sample and coenzyme (NAD+ or NADP+) were injected into the flow line by an open sandwich method. d-Malate was selectively oxidized by d-MDH when NAD+ was injected with a sample. When NADP+ was injected with a sample, l-malate was oxidized only by l-MDH. NADH or NADPH produced by the immobilized-enzyme reactors was monitored fluorometrically at 455 nm (excitation at 340 nm). Linear relationships between the responses and concentrations of d-malate and l-malate were observed in the ranges of 1 à 10â6-1 à 10â4 M and 1 à 10â6-2 à 10â4 M, respectively. The relative standard deviations for ten successive injections were less than 2% at the 0.1 mM level. This analytical method was applied to the sequential quantification of d-malate and l-malate in fruit juices and soft drinks, and the results showed good agreement with those obtained using conventional method (F-kit method).