Article ID Journal Published Year Pages File Type
10560224 Talanta 2005 6 Pages PDF
Abstract
Solid-phase extraction (SPE) procedures for cleanup and preconcentration followed by HPLC-UV method were investigated for the simultaneous determination of seven low-dosed pesticides in saline concentrates for hemodialysis. The target compounds were ametryn, desmetryn, prometryn, terbutryn, molinate, triallate and butylate. Polyethylene (three different types), teflon, polyurethane and polystyrene, in powder form, were investigated as adsorbents for solid-phase extraction of the analytes from the saline samples. Quantification was performed at 222 nm and the analytes were separated on a LiChrosorb RP-18 (5 μm, 125 mm × 4 mm i.d.) column using gradient elution with water/acetonitrile as mobile phase. The duration each chromatographic run was 18 min including column reconditioning. The efficiency of the different SPE substrates for retaining the analytes from the highly concentrated saline (HCS) samples was discussed. The best performance was achieved with polystyrene as SPE material considering preconcentration factor, precolumn clogging, reusing capability and similarity between the mobile phases for SPE and HPLC procedures. Analyte concentrations as low as 1 μg L−1 could be determined in spiked HCS samples after preconcentration on polystyrene SPE precolumns. Recoveries between 98.7 and 102.2% were obtained from commercial spiked samples. Detection limits ranging from 4.8 (for prometryn) to 46 μg L−1 (for butylate) were calculated (without preconcentration). The within-day relative standard deviations (n = 9) ranged from 2.3 to 4.8%.
Keywords
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , ,