Article ID Journal Published Year Pages File Type
10560544 Talanta 2011 7 Pages PDF
Abstract
Contamination of the exterior surface of vials of cytostatic drugs by the drugs themselves is a potential hazard to human health. This study developed a validated method using liquid chromatography-electrospray ionization tandem mass spectrometry (LC-ESI-MS/MS) for the determination of contamination of the exteriors of vials of cisplatin and carboplatin. Large Alpha® sampling swabs were employed to wipe the vial exterior. Cisplatin or carboplatin and gold(III) as an internal standard were derivatized by N,N-diethyldithiocarbamate (DDTC). Pt(DDTC)3+ and Au(DDTC)2+ were monitored by the respective transitions of m/z 639.3-490.9 and 493.0-345.0, respectively. Each separation was completed within 9 min using a 3 μm particle ODS-column. Calibration curves for cisplatin and carboplatin were linear over concentration ranges of 30-10,000 and 30-30,000 pg vial−1, respectively. The accuracies and precisions were 96.1-102.5% and within 8.2% for intra-assay and 99.6-103.3% and within 7.6% for inter-assay, respectively. Their lower limit of quantification was 30 pg vial−1. Amounts of 0.17-17.0 ng vial−1 as cisplatin and 0.48-794 ng vial−1 as carboplatin were detected from the exterior surface of the vials. This validated method using LC-ESI-MS/MS for the determination of platinum anticancer drugs is helpful for monitoring contamination of the exterior surface of drug vials.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , , ,