Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
10560544 | Talanta | 2011 | 7 Pages |
Abstract
Contamination of the exterior surface of vials of cytostatic drugs by the drugs themselves is a potential hazard to human health. This study developed a validated method using liquid chromatography-electrospray ionization tandem mass spectrometry (LC-ESI-MS/MS) for the determination of contamination of the exteriors of vials of cisplatin and carboplatin. Large Alpha® sampling swabs were employed to wipe the vial exterior. Cisplatin or carboplatin and gold(III) as an internal standard were derivatized by N,N-diethyldithiocarbamate (DDTC). Pt(DDTC)3+ and Au(DDTC)2+ were monitored by the respective transitions of m/z 639.3-490.9 and 493.0-345.0, respectively. Each separation was completed within 9 min using a 3 μm particle ODS-column. Calibration curves for cisplatin and carboplatin were linear over concentration ranges of 30-10,000 and 30-30,000 pg vialâ1, respectively. The accuracies and precisions were 96.1-102.5% and within 8.2% for intra-assay and 99.6-103.3% and within 7.6% for inter-assay, respectively. Their lower limit of quantification was 30 pg vialâ1. Amounts of 0.17-17.0 ng vialâ1 as cisplatin and 0.48-794 ng vialâ1 as carboplatin were detected from the exterior surface of the vials. This validated method using LC-ESI-MS/MS for the determination of platinum anticancer drugs is helpful for monitoring contamination of the exterior surface of drug vials.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Takashi Osawa, Takafumi Naito, Naoya Suzuki, Kimie Imai, Kunio Nakanishi, Junichi Kawakami,