Abstract
A rapid and sensitive method for the on-line separation and pre-concentration of inorganic arsenic in water samples is described. The analyte in the pentavalent oxidation state is reduced to its trivalent form with l-cysteine and the total inorganic arsenic is sorbed onto activated alumina in the acid form in a mini-column coupled to a FI-HG AAS system. Afterwards, it is eluted with 3 mol l−1 HCl. An enrichment factor of 7 was obtained, allowing an analytical flow rate of about 28 determinations per hour. The limits of detection (3σ) and of quantification (10σ) were calculated as LOD = 0.15 μg l−1 of As and LOQ = 0.5 μg l−1 of As, respectively. Relative standard deviations (n = 10) less than 8% were obtained for different arsenic concentrations and the accuracy was verified by analysing certified reference materials. Different kinds of samples, such as mineral water, drinking water, river water and natural spring water were analyzed and good agreement was obtained with the values from spiked experiments.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, ,