Article ID Journal Published Year Pages File Type
10561142 Talanta 2005 7 Pages PDF
Abstract
Determination of cadmium in urine by ETAAS suffers from severe interferences deteriorating the precision and accuracy of the analysis. Electrodeposition step prior to ETAAS allows to avoid interferences and makes cadmium determination possible even at ultratrace levels. The proposed procedures involve electrolytic deposition of cadmium from acidified urine on previously electrolytically deposited palladium film on a graphite atomizer tube, followed by removal of residual solution, pyrolysis and atomization. Both electrodeposition processes take place in a drop of the respective solution (palladium nitrate modifier and acidified urine, respectively), when Pt/Ir dosing capillary serves as an anode and the graphite tube represents a cathode. The voltage is held at −3.0 V. Matrix removal is then accomplished by withdrawal of the depleted sample solution from the tube (procedure A) or the same but followed by rinsing of the deposit with 0.2 mol l−1 HNO3 (procedure B). The accuracy of both procedures was verified by recovery test. Detection limits 0.025 and 0.030 μg Cd/l of urine were achieved for A and B procedures, respectively. Both procedures are time consuming. The measurement cycle represents 5 and 7 min for A and B procedures, respectively.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry