Article ID Journal Published Year Pages File Type
10561432 Talanta 2005 5 Pages PDF
Abstract
A hydride generation method for the determination of traces of selenium at ng mL−1 concentration ranges has been introduced using a solid mixture of tartaric acid and sodium tetrahydroborate. Atomic absorption spectrometry (AAS) has been used as the detection system. Several parameters such as the ratio of tartaric acid to sodium tetrahydroborate, type and amount of acid, and the reaction temperature were optimized by using 640 ng mL−1 (16 ng per 25 μL) of Se(IV) standard solution. The calibration curve was linear from 20 to 1200 ng mL−1 (0.5-30 ng Se(IV) per 25 μL). The relative standard deviation (%R.S.D.) of the determination was 1.93% and the detection limit was 10.6 ng mL−1 (265 pg per 25 μL) of Se(IV). The reliability of the method was checked using different types of environmental samples, such as several types of water, a sample of soil and also in a kind of calcium phosphate sample by standard addition method. For conversion of Se(VI) present in real samples to Se(IV), l-cysteine was added to NaBH4 and tartaric acid mixture. The results showed good agreement between this method and other hydride generation techniques.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , ,