A series of approaches have been assayed for FTIR determination of Mancozeb in several solid commercial fungicides using different calibration strategies. The simplest procedure was based on the use of the ratio between the absorbance of a characteristic band of Mancozeb and that of a KSCN internal standard measured in the FTIR spectra obtained from KBr pellets. It was employed the quotient between peak height absorbance values at 1525 cm−1 for Mancozeb and 2070 cm−1 for KSCN. In these conditions a precision as relative standard deviation (RSD) of 0.6% and a relative accuracy error of 0.8% (w/w) were found. For complex formulations, containing other compounds with characteristic absorption bands at different wavenumbers than Mancozeb, one of them was used as internal reference being employed the standard addition approach. In this case, the Mancozeb bands at 1525 cm−1 or at 1289 cm−1 were employed, being used the ferrocyanide band at 2075 cm−1 as internal reference. RSD values between 0.7-1.4% and a relative accuracy error of 3% (w/w) were found. A third strategy was based on the use of partial least squares (PLS) calibration. A reference set was prepared mixing Mancozeb, Kaolin, Cymoxanil and KBr, being predicted the Mancozeb concentration in pesticide formulations by using the quotient between absorbance bands of Mancozeb and those of Cymoxanil. In these conditions a relative accuracy error of 0.6% (w/w) and a relative standard deviation of 1.3% were found.