Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
10690453 | Ultrasonics | 2014 | 9 Pages |
Abstract
The UTET power transfer was carried out at a frequency of 765Â kHz, continuous wave (CW) mode. The backward data transfer was attained by inserting a 9% load resistance variation around its matched value (550Â Ohm), resulting in a 12% increase in the acoustic reflection coefficient. A backward data transmission rate of 1200Â bits/s was experimentally demonstrated using amplitude shift keying, simultaneously with an acoustic power transfer of 20Â mW to the implant.
Keywords
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Acoustics and Ultrasonics
Authors
Shaul Ozeri, Doron Shmilovitz,