Article ID Journal Published Year Pages File Type
10691017 Ultrasound in Medicine & Biology 2016 9 Pages PDF
Abstract
This prospective study evaluates the accuracy of virtual touch imaging quantification (VTIQ), a non-invasive shear wave elastography method for measuring cervical lymph nodes (LN) stiffness in differentiating benign from malignant LN. The study evaluated 270 LN in 236 patients with both conventional B-mode ultrasound and VTIQ shear wave elastography before fine-needle aspiration biopsy (FNAB). LN stiffness was measured as shear wave velocity (SWV) in m/s. Surgical resection was advised for FNAB results that were not clearly benign. Surgical pathology confirmed 54 malignant LN. The receiver operating curve (ROC) identified a single cut-off value of 2.93 m/s as the maximum SWV for predicting a malignant cervical LN. The sensitivity and specificity were 92.59% and 75.46%, respectively. Positive predictive value (PPV) was 48.54% and negative predictive value (NPV) was 97.60%. LN stiffness measured by VTIQ-generated shear wave elastography is an independent predictor of malignancy.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Acoustics and Ultrasonics
Authors
, , , , , , ,