Article ID Journal Published Year Pages File Type
10691209 Ultrasound in Medicine & Biology 2015 12 Pages PDF
Abstract
Phacoemulsification, a common treatment for cataract, is accompanied by cell damage at the corneal endothelium. Thermal exposure during treatment has been considered a reason for this damage, but a thorough experimental and theoretical assessment of the local temperature distribution inside the eye had not yet been conducted. Measurements in porcine eyes and a finite-element simulation enabled such an assessment and localized the highest temperature rise very close to the probe. The results described in this study indicate that a distance of 1 mm between the probe and the endothelium should be maintained during treatment as a safety margin, especially when fluid flow is blocked. The highest measured temperature rise with surgically reasonable system settings and unblocked fluid flow was 1.11°C. The finite-element simulation described here is able to calculate the temperature rise at the endothelium and could serve as a tool for comparing arbitrary surgical situations with respect to thermal exposure of the endothelium.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Acoustics and Ultrasonics