Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
10691371 | Ultrasound in Medicine & Biology | 2015 | 15 Pages |
Abstract
Acoustic radiation force impulse imaging and shear wave elasticity imaging (SWEI) use the dynamic response of tissue to impulsive mechanical stimulus to characterize local elasticity. A variant of conventional, multiple-track-location SWEI, denoted single-track-location SWEI, offers the promise of creating speckle-free shear wave images. This work compares the three imaging modalities using a high push and track beam density combined acquisition sequence to image inclusions of different sizes and contrasts. Single-track-location SWEI is found to have a significantly higher contrast-to-noise ratio than multiple-track-location SWEI, allowing for operation at higher resolution. Acoustic radiation force impulse imaging and single-track-location SWEI perform similarly in the larger inclusions, with single-track-location SWEI providing better visualization of small targets â¤2.5 mm in diameter. The processing of each modality introduces different trade-offs between smoothness and resolution of edges and structures; these are discussed in detail.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Acoustics and Ultrasonics
Authors
Peter J. Hollender, Stephen J. Rosenzweig, Kathryn R. Nightingale, Gregg E. Trahey,