Article ID Journal Published Year Pages File Type
10716082 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2005 5 Pages PDF
Abstract
Using simulation methods, distortion effects on energy spectra caused by errors in the energy determination have been investigated. For cosmic ray proton spectra falling steeply with kinetic energy E as E−2.7, significant effects appear. When magnetic spectrometers are used to determine the energy, the relative error increases linearly with the energy and distortions with a sinusoidal form appear starting at an energy that depends significantly on the error distribution but at an energy lower than that corresponding to the maximum detectable rigidity of the spectrometer. The effect should be taken into consideration when comparing data from different experiments, often having different error distributions.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, ,