Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
10716511 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2005 | 9 Pages |
Abstract
This paper presents a Hall probe mapping system for measuring a cyclotron magnet, which has been fabricated for the 13 MeV cyclotron at the Korea Institute of Radiological and Medical Sciences. Two Hall probes are mounted on a precision mechanical x-y stage and map magnetic field in the Cartesian coordinate system. The mapping system uses the 'flying' mode field mapping method to reduce data-acquisition time. The time required for mapping the whole gap-area of the cyclotron magnet is â¼60 min. The relative measurement error for the averaged magnetic field along beam orbit is less than 0.02%. The cyclotron magnet has been corrected using field measurement data, and the achieved total phase excursion of the cyclotron after correction is less than ±15°, which is within the tolerance of ±20° for the total phase excursion.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
K.H. Park, Y.G. Jung, D.E. Kim, B.K. Kang, M. Yoon, J.S. Chai, Y.S. Kim,