| Article ID | Journal | Published Year | Pages | File Type | 
|---|---|---|---|---|
| 10716519 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2005 | 5 Pages | 
Abstract
												The silicon strip detector (128 strips with 100 μm strip pitch) was mounted at the arm of laboratory X-ray diffractometer with 230 mm instrument radius. With the use of a LaB6 standard sample we demonstrate the angular resolution of 0.05° at low angles for classical Bragg-Brentano configuration. Experiments with a large LiF single crystal makes it possible to experimentally determine the detector limit range of linearity (30 000 cps/strip) and maximum counting range (100 000 cps/strip).
											Related Topics
												
													Physical Sciences and Engineering
													Physics and Astronomy
													Instrumentation
												
											Authors
												Jacek SÅowik, Andrzej ZiÄba, 
											