												Different codes are used for Monte Carlo (MC) calculations in radiation therapy. In this research, MCNP4C and GEANT3 codes have been compared in calculations of dosimetric characteristics of Varian Clinac 2300C/D. The parameters of influence in the differences seen in dosimetric features were discussed. This study emphasizes that both MCNP4C and GEANT3 MC can be used in radiation therapy computations and their differences in photon spectra calculations have a negligible effect on percentage depth dose computations in radiation therapy.
											Keywords
												
											Related Topics
												
													Physical Sciences and Engineering
													Physics and Astronomy
													Radiation
												
											Authors
												Asghar Mesbahi, Michael Fix, Mahmoud Allahverdi, Ellen Grein, Hossein Garaati, 
											