Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
10730486 | Applied Radiation and Isotopes | 2011 | 4 Pages |
Abstract
We investigated the ferrous sulphate-benzoic acid-xylenol orange (FBX) aqueous chemical dosimeter for measurement of dosimetric parameters such as the output factor, backscatter factor and lateral beam profiles for different square fields sizes for 60Co γ-rays. A water phantom was employed to measure these parameters. An ionization chamber (IC) was used for calibration and comparison. A comparison of the resulting measurements with an ionization chamber's measured parameters showed good agreement. We thus believe that the tissue equivalent FBX dosimetry system can measure the dosimetric parameters for 60Co with reasonable accuracy.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
O. Moussous, T. Medjadj, M. Benguerba,